Origin |
China |
Model Number |
GY-8100 |
Brand Name |
Guangyuan |
l Display model: 640*480 dot matrix LCD display with big color screen; the screen can simultaneously show many data, such as: setting values and real-time data of the power, temperature, impedance and time; frequency and time of the discharge, control mode, discharge states, instrument working states and words of alarm information.
l Operation model: PVC membrane button control.
l Alarm: Audible alarm, light alarm, information prompt display.
âÂÂÂÂ
When some faults happen to power source and power circuit of the output part
âÂÂÂÂ
When ablation stops automatically because of time over
âÂÂÂÂ
During the short circuit and open circuit happen to ablation
âÂÂÂÂ
When the protective power of the risk output is too high
The alarm prompt function would work as: discharge prompt, top limit warning, termination warning, faults warning etc..
Port Function:
âÂÂÂÂ
The port connecting with cold saline pump: cold saline can be pumped.
âÂÂÂÂ
The port connecting with the multi-channel communication.
D, 2, 3, 4---- intracardiac electronic signals can be transmitted, and can be displayed on the EP Recording System.
Serial port----the EP Recording System can control the generator, and shows following parameters: temperature, power, impedance curves.
Remote operation: can manipulate the RF generator outdoor.
Catheter port: it can be recognized automatically that non temperature-controlled catheters, thermosensitive catheters, and thermocouple catheters. Such as:
Cordis Catheter
IBI Catheter
Bard Catheter
EPT Catheter
Huitai Catheter
And other domestic Catheters
The port of back plate: FG output.
Functions of Keystroke
Self-Test after power on.
RF ENERGY: The discharge switch ensures RF energy disconnects immediately.
ENTER: Data storage function can store the current value setting. No matter power off or power on, it is not necessary to reset those values, and can accumulate frequencies and time of discharge.
MODE: According to temperature-controlled cathetersï¼ÂÂtwo modes are setï¼ÂÂnamely temperature-controlled / power controlled mode.
IMFEDANCE LIMIT: The range of impedance is manually operated for 30~300; wider inspection range.
F1, F2: Under non-discharge state, it inspects real-time impedance and temperature with the temperature-controlled catheter.
Technical Parameters and Indexes:
1. Rated Parameters âÂÂÂÂ
Output power: adjustable in the range of 2W ~ 100W continuously; the step length is 1W; the error range is ±15% in 10W ~ 100W ; the absolute error is no more than 1.5W when it is lower than 10W.
âÂÂÂÂ
Load impedance: it is 30Ω ~ 300Ω; the step length is 1Ω; the error range is ± 15%. âÂÂÂÂ
Control temperature: it is 40 â ~ 100 âÂÂÂÂ; the step length is 1âÂÂÂÂ; the error range is ± 2 âÂÂÂÂ. âÂÂÂÂ
The time and frequency of ablations: the longest time of accumulative total discharge is 999s; the error range is ± 2%; the most frequency of accumulative total discharge is 99 times.
âÂÂÂÂ
The main carrier frequencies: 450KHz; the error range is ± 20%. âÂÂÂÂ
Rated input power: 500VA +10%. 2. Security Features: âÂÂÂÂ
The RF generator should has no ablation output, on condition that it's re-connected power after power-off and stopping ablation.
âÂÂÂÂ
The equipment meets the requirements of related items of registered products standards, GB/9706.4-1999 and GB/9706.1-2007.